RESEARCH ARTICLE
Melinda Ronca-Battista1*, Pam Warkentin2, Jens Jensen3 and José - Luis Gutiérrez Villanueva4
1Tribal Air Monitoring Support Center, Institute for Tribal Environmental Professionals, Flagstaff, AZ, USA; 2Canadian National Radon Proficiency Program, Winnipeg, Canada; 3Swedish Radiation Safety Authority, Stockholm, Sweden; 4Radonova Laboratories AB, Uppsala, Sweden
The Coalition of International Radon Associations (COIRA) organised an inter-comparison of Rn-222 (radon) activity concentrations reported by calibration laboratories. A set of three AlphaGUARDs were used as transfer reference instruments against which to compare reported Rn-222 activity concentrations. Rn-222 activity concentration calibration facilities (sometimes termed chambers) from seven countries (Australia, US, Czech Republic, Spain, England, Sweden, Canada), and three continents participated in this project. The objective of the study was to provide information useful to calibration chamber operators and public health officials in the improvement of measurement and control systems, the maintenance of performance standards for measurements, and regulatory requirements for calibrations. This work builds upon and expands previous interlaboratory comparisons and provides data for estimating and using calibration uncertainty values as part of overall field error estimates and limits. A simple proportional difference between laboratories is presented here, calculated as the average for N hours of each hour’s difference between the laboratory’s concentration and the average of the three reference instruments. This percent difference ranged from less than 0.5% to just less than 8%. This work demonstrates that the ANSI/AARST standards limit of 8% for the estimated unexpanded (one sigma) individual calibration estimated uncertainty for continuous radon monitor calibration facilities in the US is achievable. However, given the few standards regarding calibration of Rn-222 activity concentration measurement instruments that themselves are often used to calibrate secondary and tertiary Rn-222 calibration facilities, there is a great need for continued interlaboratory comparisons to harmonise and document the calibration of Rn-222 activity concentrations.
Keywords: radon; radon chamber; radon reference levels; proficiency test; radon calibration; interlaboratory comparison test
Citation: Journal of the European Radon Association 2024, 5: 10607 http://dx.doi.org/10.35815/radon.v5.10607
Copyright: Journal of the European Radon Association 2024. © 2024 Melinda Ronca-Battista et al. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://creativecommons.org/licenses/by-nc/4.0/), permitting all non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Published: 12 November 2024
Competing interests and funding: The authors declare no potential conflicts of interest. The authors have not received any funding or benefits from industry or elsewhere to conduct this study.
*Melinda Ronca-Battista, Tribal Air Monitoring Support Center, Institute for Tribal Environmental Professionals, PO Box 15004, Flagstaff, AZ, USA
The Consortium of International Radon Associations (COIRA) was founded in April 2015 by a group of international radon measurement and mitigation associations who saw benefits in establishing a collaborative community for sharing technical and policy information and collaborative projects. One of the first projects of COIRA was to continue the practice of assessing and publishing the degree of consistency of reference Rn-222 activity concentrations in air reported by calibration facilities from our respective countries. Such projects are vital to assist the calibration facilities improve their understanding of their own measurement and control systems as well as their various mechanisms for generating and controlling the reference radon activity concentration in air. In addition, regulatory programs for measuring radon activity concentration in air and mitigation assume that there is some limit on measurement uncertainty, and such combined field uncertainty of radon measurements has as its minimum the initial calibration uncertainty associated with the estimated uncertainty associated with the standard reference Rn-222 activity concentration.
In the US, the commercial radon measurement industry obtains calibrations from a network of secondary and tertiary radon calibration facilities. The three secondary facilities in the US (all participants in this study) regularly compare Rn-222 activity concentrations using scintillation cells with the US primary authority, the US EPA National Analytical Radiation Environmental Laboratory. In most cases, US tertiary calibration facilities are used for in-house manufacturer research and development and interlaboratory comparisons with US secondary calibration facilities for a traceable chain of authority and documentation for their radon activity concentration in air measurement and reporting systems. These chambers and radon measurement and mitigation providers in the US participate in a voluntary accreditation program (the AARST National Radon Proficiency Program (1)), which includes policies to adhere to the ANSI/AARST standards for measurement performance (2) and quality assurance procedures for different measurement method categories (3). These standards provide guidance that, when followed, theoretically limits the resulting field uncertainty to 25% of known Rn-222 activity concentrations in air for a 48 h duration with any measurement method, based on a sequence of assumptions that are tested in periodic performance tests (1) required as a condition for continued certification. An irreducible component of estimated field error is the initial calibration error (estimated and reported as uncertainty), and the ANSI/AARST standard for Rn-222 measurement quality assurance practices requires a one sigma (unexpanded) uncertainty limit of 8% for the estimated calibration uncertainty. One method for estimating calibration uncertainty was applied to generate estimates and is used as a model for such estimates in other calibration facilities (4).
This study builds upon important previous work (5), which included 15 calibration facilities from European countries. The MetroRADON project evaluated the consistency of reference Rn-222 activity concentrations in air of 400, 1,000, and 6,000 Bq.m-3 as reported by the calibration facilities and as measured by the transfer instrument. The results showed a good consistency among all participants and a standard uncertainty of about 3% for the lower concentration range. The conclusions of the MetroRADON program included emphasis on the need for the international radon measurement community to perform this type of interlaboratory comparison on a regular basis, and for standards addressing interlaboratory comparison protocols.
The COIRA designed this study to include three continuous radon instruments together in a case with short-term alpha track detectors as a Reference Instrument Package that was shipped from calibration facility to calibration facility. Although this study provides a wealth of data, this analysis is limited to the simple presentation of the relative performance of each calibration facility, using the hourly mean of the three Rn-222 activity concentrations as the hourly benchmark, and the mean over all the hours of the exposure as the percent difference from the benchmark. The AlphaGUARD DF2000 was chosen as the transfer instrument because of the demonstrated and published sensitive response, stability, proven ruggedness, and generous loan policy of the manufacturer, and does not imply that other devices could not be used in future. The same device model was selected for all three reference devices, but our recommendation for future projects would be to repeat this project with other types/models of radon instruments in addition to this model.
The COIRA Interlaboratory Comparison Committee invited calibration facilities around the world to participate. Each facility was sent a copy of the scope of work, an agreement, and a link to a questionnaire regarding calibration facilities’ descriptions. The information obtained included chamber specifications, signed agreements regarding the study, and dates for exposure at their facilities. In total, 14 commercial, national and research calibration facilities from seven countries (Australia, US, Czech Republic, Spain, England, Sweden, Canada), and three continents participated in this project, with several facilities including multiple concentration exposures and chambers. The participating calibration facilities and the specifications provided are listed in Table 1. Information was provided by the facility and reported as provided, including images of several participating facilities in Fig. 1.
Fig. 1. Images from participating radon chambers.
The Reference Instrument Package included three AlphaGUARD DF2000 radon instruments from Bertin GmbH and a set of short-term Rapidos® alpha track detectors from Radonova Laboratories (AB) Sweden for use as transit detectors, as shown in Fig. 2. Although it would have been extremely surprising to identify exposure to high radon concentrations during shipping, and even more unexpected if this had any effect on the calibration facility exposures, the transit detectors were included to identify any significant radon exposures of the Reference Instrument Package other than in the calibration facilities’ intended exposures.
Fig. 2. Reference instrument package.
The calibration facilities were scheduled by continent so that the Reference Instrument Package was shipped between all the participating calibration facilities in one country before travelling to the next, and then to the next continent. The calibration facility’s operating costs during the exposure, and shipping costs to the next facility, were paid by each facility, so that the cost of the project was shared among participants. Additional administrative costs were paid by participating national and international radon associations. The AlphaGUARD reference devices and protective shipping box were generously provided by Bertin GmbH, and Radonova Laboratories AB provided the Rapidos® alpha track detectors.
Each calibration facility received the Reference Instrument Package with instructions to verify and note the satisfactory conditions of the equipment and the date/time, seal the alpha track detectors that were exposed during that segment of the journey, and ship them to Radonova for analysis. The instructions for exposing the three transfer reference AlphaGUARDs included:
At the end of the exposure period, calibration facilities sent their recorded hourly values for the radon activity concentration, temperature, and relative humidity from the chamber’s (own reference) instrumentation, and the AlphaGUARD output data files (*.upf2) to the project committee via email. Furthermore, and as described in the project standard operating procedures, the calibration facility opened a set of three new Rapidos® alpha track detectors, noted date/time, and placed these in the Reference Instrument Package, which were then included in the shipment to the next calibration facility.
After all the participating calibration facilities had completed their measurements, the devices were shipped to a member of the project committee for additional exposures and analysis, where the devices were compared again in a very low concentration (outdoor) radon environment.
After all the data were received from the calibration facilities, the project committee prepared the data for analysis and interlaboratory comparison and created a private online link to individual folders containing all materials and data provided by each facility. This allowed the officials from each facility to review only their original and summarised data, verify assumptions about time zones, date/times, and most importantly, confirm the exclusion of initial hours, if any, during instrument equilibration in each chamber. All participating calibration facilities were invited to two online meetings where interim results and progress reports were presented, and calibration facilities were able to ask questions or identify any clarifications regarding their data or its interpretation. All original data, assumptions, and summary statistics were available and reviewed by each calibration facility and were consistent with the interpretations of the project committee.
This project was designed to provide transparency and consistency for the participants, so included written procedures and chain of custody documentation. Upon receipt at the calibration facility, the chamber staff was responsible for the handling and data acquisition from each instrument, following the procedures and documenting the exposure on forms which were photographed and sent via email, with original paper custody forms in the shipping container, to the COIRA Committee after completion of the chamber exposures and shipping. Data and treatment protocols (6) were available, reviewed, and consistent between all participants.
In addition to standardising the procedures as much as possible, the project included measures to secure, track, and verify measurement data, as well as post study exposures to verify instrument stability and lack of accrued background during the duration of the interlaboratory comparison. These procedures and assessments included the following:
Although this study produced a wealth of data, this initial analysis only compares the response of each chamber to the response of the three transfer reference instrument (equations 1–3), evaluates the results as a function of concentration, and recommends expanding future interlaboratory comparisons.
The mean of the three transfer reference instruments for each hour is represented by rh, and individual transfer reference instruments results are ri, as shown in Equation 1.
The proportional difference between the transfer reference instruments’ hourly mean and the laboratories reported hourly mean is given by a defined proportional differences dh shown in Equation 2, where lh represents the Rn-222 activity concentration in air reported by the laboratory for each hour.
dh therefore represents the difference between the laboratory and the reference instrument package, as a proportion of the reference instrument package’s concentration, for each hour.
The statistic parameter D, given in Equation 3, represents the mean of the hourly proportional differences dh, where N is the total number of hours of the exposure.
This simple comparison statistic of the parameter D is presented in Table 2 with pertinent characteristics of each exposure. The D parameter for each exposure is represented in Fig. 3.
Fig. 3. Proportional difference D between chamber and reference instrument package with standard deviation of hourly D values as error bars.
The proportional difference (D) values for each exposure are shown in Table 2.
This analysis also included evaluating the relative responses between the three transfer reference instruments during the duration of the study and conducting final interlaboratory comparisons collocated with a fourth AlphaGUARD that had been calibrated at a commercial US facility (10). This fourth instrument was collocated with the three AlphaGUARDs in the Reference Instrument Package to provide more data at very low environments and is represented by USEPA-BMI in the Table and Figures. All the exposure data was evaluated to determine whether there was evidence of: 1) nonlinear instrument response to concentration, 2) a change over time in relative response between the individual reference instruments during the study, and 3) an effect due to background accrued during the study. All data is presented for further evaluation and with limitations described.
Figure 4 presents D as a function of the mean chamber concentration during the exposure as a representation of possible (non-linear) effect of concentration on the statistic D. As can be seen, the evidence does not indicate an effect of concentration on D, with a near zero slope. (slope with all values is 1.5E-7, and the slope without the highest concentration exposure is 1.6E-7.) Error bars represent the sample standard deviation of each D value.
Fig. 4. Mean proportional difference D as a function of mean Rn-222 activity concentration.
This analysis included a basic assessment of whether individual transfer reference instruments response functions changed relative to one another over time. We compared individual transfer reference instrument responses relative to reported chamber concentrations for the entire duration of each exposure to obtain a statistic for the individual transfer reference instrument result. The Reference Instrument Package included AlphaGUARDs with serial numbers 105, 106, and 107, and each is represented in Fig. 5. This Figure presents individual statistics for each transfer reference instrument in an attempt to reveal any consistent change between exposures in the relative responses of each transfer reference instrument.
Fig. 5. D statistics for each individual transfer reference instrument.
As can be seen in Fig. 5, while there may be a lack of homogeneity or other problems in one of the chambers (F), the average of the three reference AlphaGUARDs in that chamber (F) still resulted in a proportional difference D within the range of the other chambers, as shown in Table 2.
Additional evaluations of these data are recommended, including an assessment of bias between the three transfer reference instruments. However, for the purposes of establishing a stable benchmark against which to compare between laboratories, these data exhibited no significant change in relative reference instrument response during the study. (An additional conclusion is the recommendation that this stability and interlaboratory comparison procedures be conducted at the beginning as well as the conclusion of any future interlaboratory comparison project.)
After participant exposures were concluded, a fourth AlphaGUARD was incorporated into the project, loaned from the Tribal Air Monitoring Support Center (11) which offers within-calibration equipment loans for US federally recognized Tribal Nations. This identical DF2000 model was calibrated (independently under a contract by the US EPA) by one of the calibration facilities in the project but does provide some information on the comparability of the final responses of the three transfer reference instruments. Three sets of exposures were conducted in an outdoor environment (tent), within which all four AlphaGUARDs were collocated within an encasement of three layers of 0.1 mm plastic to eliminate the possibility of dust contamination. The results of these 2-day exposures are shown in Fig. 6.
Fig. 6. Low Rn-222 activity concentrations measured with the three reference instruments and a fourth collocated AlphaGUARD, May 2022.
The three transfer reference instruments performed well in the low concentration environment, demonstrating no consistent bias between instruments in the low Rn-222 activity concentration, as shown in Fig. 6.
Reference atmospheres are used to calibrate Rn-222 activity concentration measurement systems that establish the validity of hundreds of thousands of measurements and public and individual health decisions. This study, as well as future (and prior) interlaboratory comparisons are vital to develop a shared understanding of operations of stable Rn-222 activity concentration reference atmospheres and for the harmonisation of calibration measurement systems. Continued estimations of the consistency between different reference atmospheres provide data used to improve calibrations and to assure public health officials that mitigation decisions are based on measurement methods with limits on uncertainty that are both theoretically calculated and estimated from actual interlaboratory comparisons such as the one reported here. Protocols and quality assurance requirements in various jurisdictions in the US are based upon limiting field measurement error, a fundamental component of which is the original calibration uncertainty, as estimated by the calibrating authority and reported as calibration uncertainty. Two standards developed by ANSI/AARST and used as the basis for proficiency testing by the US radon measurement industry trade association use a value of 8% for the unexpanded (one sigma) upper limit on calibration uncertainty reported with exposures of 48 h or longer, estimated in accordance with industry practice (4). This upper bound of 8% was an estimated limit generated by the calibration facility operators serving on the MS-QA standard committee, but the results of this COIRA study indicate that this initial limit is reasonable, and perhaps can be reduced through further collaboration on calibration normalization. While the maximum D statistic from this study is not directly comparable to the maximum deviations measured in the MetroRADON interlaboratory comparison, the present interlaboratory comparison produced results which is not inconsistent with the MetroRADON study. The present study design included additional opportunities for interferences incurred during shipping between three continents, a much larger range of chamber sizes and configurations, and lasted for many more months.
Although this project does not evaluate individual calibration uncertainty estimates, it does provide an upper estimate of the maximum deviation that one internationally recognized calibration laboratory could reasonably expect to differ from another internationally recognized calibration facility, thereby informing policy makers of the current harmonization status of Rn-222 activity concentration reference atmospheres.
Regulatory and business interests coincide in encouraging future and expanded interlaboratory comparisons. Because the results are presented anonymously and without identification, individual calibration facilities may choose to publish their results and use the information to improve accuracy or for other purposes unaffiliated with COIRA.
It is important to note that the choice of the transfer reference instrument model and the number of instruments used was not based on an exhaustive determination of the best equipment for this purpose but was opportunistic and limited to three devices due to the size of some of the calibration facilities participating in this study. More interlaboratory comparisons provide more opportunities for normalizing Rn-222 activity concentrations in air between calibration facilities, thereby enabling increased accuracy in decisions impacting individual and collective public health.
This study was possible only with the generous contributions of equipment and time provided by Bertin GmbH Technologies, Radonova Laboratories AB (Sweden), and the participating calibration facilities and their operators. The project committee members are gratefult to the years of patient work on this project committed by participants and equipment providers. We also acknowledge G. Budd, of the US EPA, who envisioned this project decades before its realisation, and James McLaughlin, University College Dublin/European Radon Association, who put forward the idea as the first COIRA initiative, and who along with Martin Freeman, UK Radon Association (in memoriam), who encouraged participation by by all radon calibration facilities early the project.
1. | National Radon Proficiency Program, American Association of Radon Scientists and Technologists. Certification of measurement providers. 2023. Available from: https://nrpp.info/devices/approved-devices/ [9 October 2024]. |
2. | Performance specifications for instrumentation systems designed to measure radon gas in air. ANSI/AARST MS-PC 2015. 2015. Available from: https://standards.aarst.org/MS-PC-2015/ [9 October 2024]. |
3. | Radon Measurement Systems Quality Assurance. ANSI/AARST MS-QA-2019. Available from: https://webstore.ansi.org/standards/aarst/ansiaarstmsqa2019 [9 October 2024]. |
4. | Ronca-Battista M, Budd G, McLemore S. Uncertainty of 222Rn concentrations in the USEPA radiation and indoor environments national laboratory exposure chamber. Health Phys 2013; 104(2): 168–78. doi: 10.1097/HP.0b013e3182712e5a |
5. | Beck TR, Antohe A, Cardellini F, Cucoş A, Fialova E, Grossi C, et al. The metrological traceability, performance and precision of European radon calibration facilities. Int J Environ Res Public Health 2021; 18(22): 12150. doi: 10.3390/ijerph182212150 |
6. | R Core Team. R: a language and environment for statistical computing. Vienna, Austria: R Foundation for Statistical Computing; 2021. Available from: https://www.R-project.org/ [9 October 2024]. |
7. | Lin CF, Wang JJ, Lin SJ, Lin CK. Performance comparison of electronic radon monitors. Appl Radiat Isot 2013; 81: 238–41. doi: 10.1016/j.apradiso.2013.03.024 |
8. | Roessler F, Buerkin W, Villert J. AlphaGUARD, the new reference for continuous radon monitoring in air, soil, gas, water and material. In Wilhelm C, ed. Radiation protection for humans and environment 50 years competence in the professional association. Germany; 2016, p. 468. Available from: https://inis.iaea.org/search/citationdownload.aspx [9 October 2024]. |
9. | Federal Office for Radiation Protection (BfS). BFS, Ionizing Radiation, Radon Calibration Laboratory. Available from: https://www.bfs.de/EN/topics/ion/environment/laboratories/radon/radon.html [9 October 2024]. |
10. | Bowser Morner International Radon Calibration Services. Radon reference laboratory services include: Calibration of continuous radon monitors 2023. Available from: https://www.bowser-morner.com/radon-reference-lab [9 October 2024]. |
11. | Tribal Air Monitoring Support Center. TEMS Services: Indoor Air Quality. Las Vegas, NV. Available from: https://www7.nau.edu/itep/main/tams/Services/EqpLoans [9 October 2024]. |